Influence of Intraocular Lens Asphericity and Blue Light Filtering on Visual Outcome, Contrast Sensitivity, and Aberrometry after Uneventful Cataract Extraction

Abstract

Purpose: To evaluate the effect of asphericity and blue light filter (BLF) of three different intraocular lenses (IOLs) on the visual performance, second- and third-order aberrations (defocus, coma, trefoil), and contrast sensitivity after uneventful cataract surgery.


Methods: One hundred and twenty eyes of 60 patients with clinically significant cataract were randomly assigned to receive one of the three IOL types: Bioline Yellow Accurate (aspheric, with BLF, i-medical, Germany), BioAcryl 60125 (spherical, without BLF, Biotech, France), and H65C/N (aspheric, without BLF, PhysIOL, Belgium). Each IOL was implanted in 40 eyes. Complete ophthalmologic examination, functional acuity contrast testing and wavefront analysis were performed 60 days postoperatively.


Results: The mean postoperative best-corrected visual acuity (BCVA) was 0.95 ± 0.08, not differing statistically among the IOL groups (P = 0.83). Mean defocus and coma values did not yield any statistically significant difference through the IOL groups varying from –0.784 to –0.614 μm and 0.129 to 0.198 μm (P = 0.79 and 0.34, respectively). Bioline Yellow Accurate IOL presented less trefoil aberrations, 0.108 ± 0.05 μm, compared to the other two IOL types (BioAcryl [0.206 ± 0.19 μm] and Physiol [0.193 ± 0.17 μm], P < 0.05). Contrast sensitivity values did not differ among the groups under all lighting conditions. Bioline Yellow IOL showed a statistically higher loss of contrast sensitivity (between mesopic and mesopic with glare conditions) compared to the BioAcryl and PhysIOL in 12 and 3 cpd spatial frequencies, respectively (P < 0.05).


Conclusion: Bioline Yellow IOL indicated lower contrast sensitivity under mesopic conditions when glare was applied but resulted in less trefoil aberrations after uneventful cataract surgery. No further differences were noted in postoperative visual performance among three IOL groups.

Keywords:

Aberrometry, Asphericity, Blue-light Filtering, Contrast Sensitivity, Intraocular Lens

References
1. Mainster MA. Violet and blue light blocking intraocular lenses: photoprotection versus photoreception. Br J Ophthalmol 2006;90:784–792.

2. Schwiegerling J. Blue-light-absorbing lenses and their effect on scotopic vision. J Cataract Refract Surg 2006;32:141–144.

3. Yuan Z, Reinach P, Yuan J. Contrast sensitivity and color vision with a yellow intraocular lens. Am J Ophthalmol 2004;138:138–140.

4. Cuthbertson F, Peirson S, Wulff K, Foster RG, Downes SM. Blue light-filtering intraocular lenses: review of potential benefits and side effects. J Cataract Refract Surg 2009;35:1281–1297.

5. Henderson B, Grimes K. Blue-blocking IOLs: a complete review of the literature. Surv Ophthalmol 2010;55:284–289.

6. Ham WT Jr, Mueller HA, Sliney DH. Retinal sensitivity to damage from short wavelength light. Nature 1976;260:153–155.

7. Tomany SC, Cruickshanks KJ, Klein R, Klein BEK, Knudtson MD. Sunlight and the 10-year incidence of age-related maculopathy: the Beaver Dam Eye study. Arch Ophthalmol 2004;122:750–757.

8. Taylor HR, West S, Munoz B, Bressler SB, Bressler NM. The long term effects of visible light on the eye. Arch Ophthalmol 1992;110:99–104.

9. Hayashi K, Hayashi H. Visual function in patients with yellow tinted intraocular lenses compared with vision in patients with non-tinted intraocular lenses. Br J Ophthalmol 2006;90:1019–1023.

10. Olson MD, Miller KM. Implanting a clear intraocular lens in one eye and a yellow lens in the other eye: a case series. Am J Ophthalmol 2006;141:957–958.

11. Olson RJ, Werner L, Mamalis N, Cionni R. New intraocular lens technology. Am J Ophthalmol 2005;140:709–716.

12. Cionni RJ, Tsai JH. Color perception with acrysof natural and acrysof single-piece intraocular lenses under photopic and mesopic conditions. J Cataract Refract Surg 2006;32:236–242.

13. Brockmann C, Schulz M, Laube T. Transmittance characteristics of ultraviolet and blue-light-filtering intraocular lenses. J Cataract Refract Surg. 2008;34:1161– 1166.

14. Applegate RA, Sarver EJ, Khemsara V. Are all aberrations equal? J Refract Surg 2002;18:556–562.

15. Tzelikis PF, Akaishi L, Trindade FC, Boteon JE. Spherical aberration and contrast sensitivity in eyes implanted with aspheric and spherical intraocular lenses: a comparative study. Am J Ophthalmol 2008;145:827–833.

16. Pandita D, Raj SM, Vasavada VA, Vasavada VA, Kazi NS, Vasavada AR. Contrast sensitivity and glare disability after implantation of acrysof IQ natural aspherical intraocular lens; prospective randomized masked clinical trial. J Cataract Refract Surg 2007;33:603–610.

17. Tzelikis PF, Akaishi L, Trindade FC, Boteon JE. Ocular aberrations and contrast sensitivity after cataract surgery with acrysof IQ intraocular lens implantation; clinical comparative study. J Cataract Refract Surg 2007;33:1918–1924.

18. Rocha KM, Soriano ES, Chalita MR, Yamada AC, Bottós K, Bottós J, et al. Wavefront analysis and contrast sensitivity of aspheric and spherical intraocular lenses: a randomized prospective study. Am J Ophthalmol 2006;142:750–756.

19. Kershner RM. Retinal image contrast and functional visual performance with aspheric, silicone, and acrylic intraocular lenses; prospective evaluation. J Cataract Refract Surg 2003;29:1684–1694.

20. Denoyer A, Le Lez M-L, Majzoub S, Pisella P-J. Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation; effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision. J Cataract Refract Surg 2007;33:210–216.

21. BellucciR, Morselli S. Optimizing higher-order aberrations with intraocular lens technology. Curr Opin Ophthalmol 2007;18:67–73.

22. Downie LE, Busija L, Keller PR. Blue-light filtering intraocular lenses (IOLs) for protecting macular health. Cochrane Database Syst Rev 2018;22:CD011977.

23. Downes SM. Ultraviolet or blue-filtering intraocular lenses: what is the evidence? Eye 2016;30:215–221.

24. Montés-Micó R, Alió JL, Muñoz G, Pérez-Santonja JJ, Charman WN. Postblink changes in total and corneal ocular aberrations. Ophthalmology 2004;111:758–767.

25. Kim SW, Ahn H, Kim EK, Kim T-I. Comparison of higher order aberrations in eyes with aspherical or spherical intraocular lenses. Eye 2008;22:1493–1498.

26. Caporossi A, Martone G, Casprini F, Rapisarda L. Prospective randomized study of clinical performance of 3 aspheric and 2 spherical intraocular lenses in 250 eyes. J Refract Surg 2007;23:639–648.

27. Kurz S, Krummenauer F, Thieme H, Dick HB. Contrast sensitivity after implantation of a spherical versus an aspherical intraocular lens in biaxial microincision cataract surgery. J Cataract Refract Surg 2007;33:393–400.

28. Kasper T, Buhren J, Kohnen T. Visual performance of aspherical and spherical intraocular lenses: intraindividual comparison of visual acuity, contrast sensitivity, and higher-order aberrations. J Cataract Refract Surg 2006;32:2022–2029.

29. Munoz G, Albarran-Diego C, Montes-Mico R, Rodríguez- Galietero A, Alió JL. Spherical aberration and contrast sensitivity after cataract surgery with the Tecnis Z9000 intraocular lens. J Cataract Refract Surg 2006;32:1320– 1327.

30. Kara-Junior N, Espindola RF, Gomes BA, Ventura B, Smadja D, et al. Effects of blue light-filtering intraocular lenses on the macula, contrast sensitivity, and color vision after a long-term follow-up. J Cataract Refract Surg 2011;37:2115–2119.

31. Greenstein VC, Chiosi F, Baker P, Seiple W, Holopigian K, et al. Scotopic Sensitivity and Color Vision with a Blue- Light-Absorbing Intraocular Lens. J Cataract Refract Surg 2007;33:667–672.

32. Muftuoglu O, Karel F, Duman R. Effect of a yellow intraocular lens on scotopic vision, glare disability, and blue color perception. J Cataract Refract Surg 2007;33:658–666.