Corneal Parameters in Healthy Subjects Assessed by Corvis ST

Abstract

Purpose: To evaluate corneal biomechanics using Corvis ST in healthy eyes from Iranian keratorefractive surgery candidates.


Methods: In this prospective consecutive observational case series, the intraocular pressure (IOP), central corneal thickness (CCT), and biomechanical properties of 1,304 eyes from 652 patients were evaluated using Corvis ST. Keratometric readings and manifest refraction were also recorded.


Results: The mean (±SD) age of participants was 28 ± 5 years, and 31.7% were male. The mean spherical equivalent refraction was –3.50 ± 1.57 diopters (D), the mean IOP was 16.8 ± 2.9 mmHg, and the mean CCT was 531 ± 31 μm for the right eye. The respective means (±SD) corneal biomechanical parameters of the right eye were as follows: first applanation time: 7.36 ± 0.39 milliseconds (ms); first applanation length: 1.82 ± 0.22 mm; velocity in: 0.12 ± 0.04 m/s; second applanation time: 20.13 ± 0.48 ms; second applanation length: 1.34 ± 0.55 mm; velocity out: –0.67 ± 0.17 m/s; total time: 16.84 ± 0.64 ms; deformation amplitude: 1.05 ± 0.10 mm; peak distance: 4.60 ± 1.01 mm; and concave radius of curvature: 7.35 ± 1.39 mm. In the linear regression analysis, IOP exhibited a statistically significant association with the first and second applanation times, total time, velocity in, peak distance, deformation amplitude, and concave radius of curvature.


Conclusion: Our study results can be used as a reference for the interpretation of Corvis ST parameters in healthy refractive surgery candidates in the Iranian population. Our results confirmed that IOP is a major determinant of Corvis parameters.

Keywords:

Central Corneal Thickness, Corneal Biomechanics, Corvis ST, Intraocular Pressure

References
1. Huseynova T, Waring GO, Roberts C, Krueger RR, Tomita M. Corneal biomechanics as a function of intraocular pressure and pachymetry by dynamic infrared signal and Scheimpflug imaging analysis in normal eyes. Am J Ophthalmol 2014;157:885–893.

2. Pinero DP, Alcon N. In vivo characterization of corneal biomechanics. J Cataract Refract Surg 2014;40:870–887.

3. Ambrósio R, Ramos I, Luz A, Faria FC, Steinmueller A, Krug M, et al. Dynamic ultra high speed Scheimpflug imaging for assessing corneal biomechanical properties. Rev Bras Oftalmol 2013;72:99–102.

4. Hassan Z, Modis L, Jr, Szalai E, Berta A, Nemeth G. Examination of ocular biomechanics with a new Scheimpflug technology after corneal refractive surgery. Cont Lens Anterior Eye 2014;37:337–341.

5. Hong J, Xu J, Wei A, Deng SX, Cui X, Yu X, et al. A new tonometer–the Corvis ST tonometer: clinical comparison with noncontact and Goldmann applanation tonometers. Invest Ophthalmol Vis Sci 2013;54:659–665.

6. Tian L, Huang YF, Wang LQ, Bai H, Wang Q, Jiang JJ, et al. Corneal biomechanical assessment using corneal visualization scheimpflug technology in keratoconic and normal eyes. J Ophthalmol 2014;2014:147516.

7. Reznicek L, Muth D, Kampik A, Neubauer AS, Hirneiss C. Evaluation of a novel Scheimpflug-based noncontact tonometer in healthy subjects and patients with ocular hypertension and glaucoma. Br J Ophthalmol 2013;97:1410–1414.

8. Vellara HR, Ali NQ, Gokul A, Turuwhenua J, Patel DV, McGhee CN. Quantitative analysis of corneal energy dissipation and corneal and orbital deformation in response to an air-pulse in healthy eyes. Invest Ophthalmol Vis Sci 2015;56:6941–6947.

9. Wang W, He M, He H, Zhang C, Jin H, Zhong X. Corneal biomechanical metrics of healthy Chinese adults using Corvis ST. Cont Lens Anterior Eye 2017;40:97–103.

10. Lee H, Kang DSY, Ha BJ, Choi JY, Kim EK, Seo KY, et al. Biomechanical properties of the cornea using a dynamic scheimpflug analyzer in healthy eyes. Yonsei Med J 2018;59:1115–1122.

11. Valbon BF, Ambrosio R, Jr, Fontes BM, Luz A, Roberts CJ, Alves MR. Ocular biomechanical metrics by Corvis ST in healthy Brazilian patients. J Refract Surg 2014;30:468–473.

12. Chua J, Tham YC, Liao J, Zheng Y, Aung T, Wong TY, et al. Ethnic differences of intraocular pressure and central corneal thickness: the Singapore Epidemiology of Eye Diseases study. Ophthalmology 2014;121:2013–2022.

13. Fern KD, Manny RE, Gwiazda J, Hyman L, Weise K, Marsh- Tootle W. Intraocular pressure and central corneal thickness in the COMET cohort. Optom Vis Sci 2012;89:1225–1234.