Changes in Pupil Area During Low-energy Femtosecond Laser Assisted Cataract Surgery

Abstract

Purpose: To study the potential changes in pupil area within low-energy femtosecond-laser assisted cataract surgery (FLACS).


Methods: A retrospective assessment of the pupil size was performed in the eyes undergoing FLACS using the Ziemer LDV Z8. We measured the pupil diameters as part of the images taken preoperatively and at the completion of laser pretreatment (after releasing the suction). We calculated the pupil area in 40 eyes of 40 patients (14 right and 26 left eyes). The mean ± standard deviation (SD) of age of the patients was 74 ± 7.4 years (range: 51-87). Paired t-test was used for statistical analyses. Subgroups were built with reference to age and preoperative pupil area (smaller than or equal to the median versus larger than the median).


Results: The mean ± SD axial length, anterior chamber depth, white-to-white distance and lens thickness were 24.01 ± 1.47, 3.23 ± 0.4, 11.97 ± 0.49, and 4.59 ± 0.41 mm, respectively. The mean ± SD pupil area was 39.33 ± 7.1 mm2 preoperatively and 39.3 ± 6.75 mm2 after laser pretreatment. The mean ± SD change in pupil area was -0.03 ± 2.12 mm2. There were no statistically significant changes between preoperative and post-laser pupil areas (P = 0.93, 95% CI: -0.71 to 0.65). Comparisons within subgroups also did not detect pupil area reduction.


Conclusion: This study did not detect statistically significant changes in pupil area after laser pretreatment using low-energy FLACS. This observation is in contrast to previous studies using other laser platforms.

Keywords:

Cataract Surgery; Femtosecond Laser; Pupil Size; Safety

References
1. Dick HB, Schultz T. A review of laser-assisted versus traditional phacoemulsification cataract surgery. Ophthalmol Ther 2017;6:7-18.

2. Ewe SY, Abell RG, Vote BJ. Femtosecond laser-assisted versus phacoemulsification for cataract extraction and
intraocular lens implantation: clinical outcomes review. Curr Opin Ophthalmol 2018;29:54-60.

3. Grewal DS, Schultz T, Basti S, Dick HB. Femtosecond laser-assisted cataract surgery–current status and future
directions. Surv Ophthalmol 2016;61:103-131.

4. Nagy Z, Takacs A, Filkorn T, Sarayba M. Initial clinical evaluation of an intraocular femtosecond laser in cataract
surgery. J Refract Surg 2009;25:1053-1060.

5. Lundstrom M, Dickman M, Henry Y, Manning S, Rosen P, Tassignon MJ, et al. Femtosecond laser-assisted cataract surgeries reported to the European registry of quality outcomes for cataract and refractive surgery: baseline characteristics, surgical procedure, and outcomes. J Cataract Refract Surg 2017;43:1549-1556.

6. Nejat F, Sarahati S, Nobari SM, Jadidi K, Naderi M, Nejat MA. Preliminary results of femtosecond laser-assisted
cataract surgery in a private clinic in Iran. J Ophthalmic Vis Res 2017;12:39-43.

7. Nagy ZZ, Takacs AI, Filkorn T, Kranitz K, Gyenes A, Juhasz E, et al. Complications of femtosecond laser-assisted
cataract surgery. J Cataract Refract Surg 2014;40:20-28.

8. Roberts TV, Lawless M, Bali SJ, Hodge C, Sutton G. Surgical outcomes and safety of femtosecond laser cataract
surgery: a prospective study of 1500 consecutive cases. Ophthalmology 2013;120:227-233.

9. Hashemi H, Seyedian MA, Mohammadpour M. Small pupil and cataract surgery. Curr Opin Ophthalmol 2015;26:3-9.

10. Bali SJ, Hodge C, Lawless M, Roberts TV, Sutton G. Early experience with the femtosecond laser for cataract
surgery. Ophthalmology 2012;119:891-899.

11. Chang JS, Chen IN, Chan WM, Ng JC, Chan VK, Law AK. Initial evaluation of a femtosecond laser system in cataract surgery. J Cataract Refract Surg 2014;40:29-36.

12. Diakonis VF, Kontadakis GA, Anagnostopoulos AG, Yesilirmak N, Waren DP, Cabot F, et al. Effects of short-term preoperative topical ketorolac on pupil diameter in eyes undergoing femtosecond laser-assisted capsulotomy. J Refract Surg 2017;33:230-234.

13. Jun JH, Hwang KY, Chang SD, Joo CK. Pupil-size alterations induced by photodisruption during femtosecond
laser-assisted cataract surgery. J Cataract Refract Surg 2015;41:278-285.

14. Diakonis VF, Yesilirmak N, Sayed-Ahmed IO, Warren DP, Kounis GA, Davis Z, et al. Effects of femtosecond
laser-assisted cataract pretreatment on pupil diameter: a comparison between three laser platforms. J Refract Surg 2016;32:84-88.

15. Jun JH, Yoo YS, Lim SA, Joo CK. Effects of topical ketorolac tromethamine 0.45% on intraoperative miosis and prostaglandin E2 release during femtosecond laser-assisted cataract surgery. J Cataract Refract Surg
2017;43:492-497.

16. Schultz T, Joachim SC, Stellbogen M, Dick HB. Prostaglandin release during femtosecond laser-assisted
cataract surgery: main inducer. J Refract Surg 2015;31:78-81.

17. Mariacher S, Ebner M, Seuthe AM, Januschowski K, Ivanescu C, Opitz N, et al. Femtosecond laser-assisted
cataract surgery: first clinical results with special regard to central corneal thickness, endothelial cell count, and aqueous flare levels. J Cataract Refract Surg. 2016;42:1151-1156.

18. Schultz T, Joachim SC, Szuler M, Stellbogen M, Dick HB. NSAID pretreatment inhibits prostaglandin release
in femtosecond laser-assisted cataract surgery. J Refract Surg 2015;31:791-794

19. Ebner M, Mariacher S, Januschowski K, Boden K, Seuthe AM, Szurman P, et al. Comparison of intraocular pressure during the application of a liquid patient interface (FEMTO LDV Z8) for femtosecond laser-assisted cataract surgery using two different vacuum levels. Brit J Ophthalmol 2017;101:1138-1142.

20. Williams GP, Ang HP, George BL, Liu YC, Peh G, Izquierdo L, et al. Comparison of intra-ocular pressure changes with liquid or flat applanation interfaces in a femtosecond laser platform. Sci Rep 2015;5:14742.

21. Schultz T, Conrad-Hengerer I, Hengerer FH, Dick HB. Intraocular pressure variation during femtosecond laserassisted cataract surgery using a fluid-filled interface. J Cataract Refract Surg 2013;39:22-27.

22. Walter K, Delwadia N, Coben J. Continuous intracameral phenylephrine-ketorolac irrigation for miosis prevention in femtosecond laser-assisted cataract surgery: reduction in surgical time and iris manipulation. J Cataract Refract Surg 2019;45:465-469.